This overview provides quick start-up steps for operating your Model 1003 Indicator.
Full details are described on the following pages.
Make all necessary wiring connections as described on the proceeding pages
Enter into Program Mode by pressing “PROG” button and press button until you see “CAL” screen
appear. Set the value to the desired end of stroke displacement (inches/degrees).
Step through each digit
using the “STEP” button and increase or decrease to the appropriate digit using the “UP” and
“DOWN” buttons
. Pressing the “PROG” button again will bring you to the “DECIMAL POINT” screen.
Adjust the decimal point as appropriate to ensure the correct calibration value is set. When complete
press the “PROG” button again to exit Program Mode.
For more information on programming values see
proceeding page 20.
Turn “CAL” and “ZERO” lights OFF by pressing corresponding buttons to activate voltmeter function
Locate NULL (or zero) position of the transducer – display will read nearly all “0”s
(This is the center of the transducer’s working range)
Displace the transducer to point in stroke where ZERO is desired.
(If you choose NULL position, you’re already there!)
Hit the “ZERO” button on front panel.
Displace the transducer to point in stroke where programmed CAL value is desired
(Be sure to stay within the working range of transducer)
Hit the “CAL” button on front panel
CALIBRATION IS NOW COMPLETE!
Summary of Contents for 1003
Page 1: ...M009 0009 Rev 8 MODEL 1003 TRANSDUCER INDICATOR INSTRUCTION MANUAL ...
Page 13: ... 13 Table 3 Table 4 Figure 3 RS232C Option Switch Settings and Pin Designations ...
Page 15: ... 15 S1 S2 S3 S4 SWITCH DEFINITIONS Front Panel View ...
Page 23: ... 23 Table 6 Troubleshooting Guide Table 7 Replacement Mating Connector P Ns ...