![Tradinco TRAQC-7 User Manual Download Page 40](http://html1.mh-extra.com/html/tradinco/traqc-7/traqc-7_user-manual_1148138040.webp)
User Manual
TRAQC-8
TRAQC-8 04/2017 EN Rev.4
page : 40
7.
Product Specifications
7.1
General
Pressure range
: See data sheet
Test connection
: Minimess 1215 quick connect coupling with internal shut off valve
Test hose/filling hose
: Minimess 1215,
¼” NPT male coupling
Filling hose
: Minimess 1215 + Ru3
Reducer
: Model 8810 max 350 bar
Venting valve
: Model T 1700 soft seat, tight shut-off.
Monitor gauge
: 3 Pieces
Gas bottles
: Maximal working pressure 200 bar, capacity 2 x 1 liter
Safety
: Gas bottles protect by safety valve, set at 220 bar
Power supply cable
: Three phases cable for euro plug, length 2.5 meter
Power supply
: 220 VAC 50/60 Hz
Carrying case
: Portable case in Aluminum.
Dimensions
: (L x W x H) 405x295x255 mm
Weight
: ± 17 kg.
Display
: LCD, 70 x 70 mm, 128 x 128 pixels, with backlight
Resolution
: 4, 5 of 6 digits
Operating temp.
: 5 - 45 º C
Certification
: CE
Interface
: RS232 Interface
Menu language
: NL, UK, FR, D
Connections
: Pressure Minimess 1215
: Electric: banana socket 4mm
: RS232
Loop supply for external powered transmitters
250 Ohm resistance for HART
®
transmitters
7.2
Measurement
Pressure, integrated sensor : accuracy 0.02% FS
Sensor Ranges
: 16, 20, 40, 60 , 100, 160 and 250 barg (others on request)
Electric
: 4mm 250 Ohm resistance for HART® transmitters
DC measurement
: 0
–
30 V, 0.02% R+ 0.01 FS, 0
–
24 mA, 0.02% R+ 0.01 FS.
Switch test
7.3
Generation and control
Pressure
: Up to 200 bar pressure from the internal gas bottles.
DC generation, Option
: 0
–
11 V, 0.02% R+ 0.01 FS, 0
–
24 mA, 0.025% R+ 0.01 FS.
7.4
Data storage
Large internal memory 2MB suitable for storage of 32 calibration procedures and calibration
results. Suitable for up- and downloading in combination with AutoCal
®