User Manual
TRAQC-8
TRAQC-8 04/2017 EN Rev.4
page : 36
3.10.2.3
Voltage output
In order to calibrate the Voltage signal, connect the ‘PLUS’ and
‘MINUS’ 4mm banana sockets of the TRAQC
-8 to a multi meter
with higher specifications than the TRAQC-8.
•
Correct the output signal by using the ‘ARROW’ keys.
When
the desired setting is reached press ‘ENTER’ key
(F1).
•
In the following screen, fill in the readout of the multi
meter by using the ‘ARROW’ keys and press the ‘ENTER’
key (F1) to confirm.
•
Repeat the same procedure for the maximum value.
3.10.3
Show calibration factors
In this menu values are stated which are used by Tradinco
Instruments for calibrating the TRAQC-8.