User Manual
TRAQC-8
TRAQC-8 04/2017 EN Rev.4
page : 18
3.
User Configuration
3.1
Pressure Setup
The “Pressure Setup” menu shows the options for pressure inputs.
After pressing the ‘MENU’
key
(F1), press the ‘ARROW’ keys until
the “pressure setup” is selected, then press ‘ENTER’
key (F1). The
available pressure options for the instrument are now displayed
(see figure) and it is now possible to select one of options with the
‘ARROW’ keys. The d
ifferent options of the pressure setup are
described in the following paragraphs.
3.1.1
Pressure range
This menu option is available in instruments with 2 or more
pressure sensors installed.
When you select this option a menu with the available pressure
ranges is shown,
You can select a line with the arrow keys and then switch the
selected pressure range on or off depending upon its current status
when the electrical ranges are not active, otherwise you can select
only one range to be displayed. See also section 2.11.
3.1.2
Pressure unit
The menu to select the pressure unit can in most operating modes be accessed by pressing the
‘UNIT’ function key, There are however operating modes where this is not possible , for these
modes the pressure unit selection is also available in the pressure setup menu, see section 2.10.
3.1.3
Averaging on/off
A progressing averaging can be applied. The reading speed is standard three times per second.
The objective is to display the fluctuating measuring signals in a reliable manner. If averaging is
switched on, this is shown by ‘
𝑥̅
’ above the pressure unit reading in the ‘measuring mode’ menu.
To enable or disable the averaging function select the “averaging on/off” option in the “pressure
setup” menu.
The default
state is the “averaging off”. To change the current state press ‘ENTER’
key (F1).
3.1.4
Averaging setup
The number of times that averaging must be performed can be set
as follows: Select the “averaging setup” in the pressure setup
menu with the ‘ARROW’ keys and press ‘ENTER’
key (F1) to
confirm. Now use the ‘ARROW’ keys to change the number of cycles
to the
desired one, press ‘ENTER’
key (F1) to confirm.
The number of cycles default is 6. The cycles can be set from 3 to
9.