34
www.trace2o.com
Metalyser HM5000 Benchtop
Instruction Manual v1.0
Cyclic Voltammetry
Cyclic Voltammetry is a reversible linear sweep. The scan starts at the potential
Ebegin and scans to Elow at which point it reverses and scans positive to
Eend before reversing again and scanning back to the start potential.
NB:
Due to technical limitations of the instrument, the cyclic voltammetry function
is limited to a maximum of 400 data points. If you are encountering problems with
the cyclic voltammetry function, ensure that Ehigh – Elow *2 /Estep < 400.
Potential
Time