5. Error Display
5.1 Error Display of Instrument
Error display
Contents
Remedial measure
Because the instrument
inside is hot, the
protective circuit works.
Turn off the power.
Turn off the power and leave the
instrument as it is under the use
condition (5 - 35°C) for about 30
minutes. Then, turn on the power.
The level of the backup
battery power is lowered.
If you continue the
operation, press the
“FIELD” switch.
The level of the measurement data
backup battery power is lowered
and the data stored in the
measuring instrument may be
erased. It is necessary to replace
the battery. Contact our company
or the retailer shop where you
purchased the instrument. When
you do not turn off the power, this
error does not have influence on
measurement.
The instrument exceeds
the measurement range.
Press any of the switches.
Take the following remedial
measures or others.
- Lower the brightness of the
measurement target.
- Select a smaller measuring angle.
- Insert a neutral density filter.
The external
synchronizing signal
capturing error occurs.
Press any of the switches.
The external synchronizing signal
is not captured normally. Check
the synchronizing signal generating
device and the connection of BNC
cable.
E005
– E007
Communication error
occurs.
Check the communication
conditions.
E900 - E999
Internal error occurs.
Press any of the switches.
Internal error occurs.
Contact our company or the retailer
shop where you purchased the
instrument.
• When the error is still displayed after you have taken the above-mentioned remedial measures,
repair may be necessary. Contact our company or the retailer shop where you purchased the
instrument.
***
ERROR
***
E9XX
XXXXXXXXXXXXX
hit
any
key
***
ERROR
***
EXXX
Comm
E
rror
***
ERROR
***
E004
Sync
Error
hit
any
key
***
ERROR
***
E001
Over
Range
hit
any
key
Attention!!
Low
Battery
Continue
:
FIELD
Attention!!
Temperature
Not
set
Please
Power
OFF