3.3.1 Setting Delay Time
When the integral time delay function is set to “ON”, set the integral time.
Note
It is recommended to set 100 cycles or more as the integral time.
Example: When measuring the light source with the cycle discrepancy 10%:
When measuring with the integral time of 10 cycles, the discrepancy is as follows.
0.1t/10t = 1% (“t” is the cycle time.)
When measuring with the integral time of 100 cycles, the discrepancy is as follows.
0.1t/100t = 0.1%
The setting procedures are described below.
Setting
1.
Shift to the function mode. Access [MEASURE]
from the function menu. Then, access the [Integ
Delay Time] screen.
☞
“3.1.2 Entering/Returning from the Function Mode”.
2.
Input the desired numerical value according to the numerical value input method. You can
input the numerical values within the “50 – 3000 [ms]” range.
Setting method:
☞
“3.1.3 Setting Numerical Value”.
3.
Press the [ENTER] switch to decide the value.