12-2
12 ERROR DISPLAYS
z
If error still persist after attempting to clear them, contact your local Topcon dealer or Topcon head
office.
TILT OVER
At the tilt ON setting, the
instrument is tilted 6' or more and
under 10'.
Accurately level the instrument.
TILT ERROR
At the tilt ON setting, the
instrument is tilted 10' or more.
SCAN AREA TOO SMALL
PLEASE START OVER
The starting point and end point
of the designated scan range are
too close together.
Designate a wider range.
INAPPROPRIATE VALUE
PLEASE INUPUT AGAIN
The entered value is out of the
limited range.
Enter a value within the limited
range.
TOO MANY H POINTS
PLEASE INUPUT AGAIN
Enter a value not exceeding the
maximum horizontal points.
TOO FEW H POINTS
PLEASE INUPUT AGAIN
Enter a value equal to or above
the minimum horizontal points.
TOO MANY V POINTS
PLEASE INUPUT AGAIN
Enter a value not exceeding the
maximum vertical points.
TOO FEW V POINTS
PLEASE INUPUT AGAIN
Enter a value equal to or above
the minimum vertical points.
TOO MANY TOTAL PTS
PLEASE INUPUT AGAIN
Enter a value not exceeding the
maximum total points
TOO FEW TOTAL PTS
PLEASE INUPUT AGAIN
Enter a value equal to or above
the minimum total points.
H INTERVAL TOO SMALL
PLEASE INUPUT AGAIN
Enter a value equal to or above
the minimum horizontal interval.
H INTERVAL TOO LARGE
PLEASE INUPUT AGAIN"
Enter a value not exceeding the
maximum horizontal interval.
V INTERVAL TOO SMALL
PLEASE INUPUT AGAIN
Enter a value equal to or above
the minimum horizontal angle.
V INTERVAL TOO LARGE
PLEASE INUPUT AGAIN
Enter a value not exceeding the
maximum horizontal angle.
INTERNAL SCAN ERROR
PLEASE START OVER
An error occurred during
scanning.
Unable to scan. Long-push the
power key and restart. If long-
pushing the power key does not
work, remove the internal and
external batteries and restart. If
the error persists, have the
instrument repaired.
BATTERY LOW LEVEL!
The battery is low. Or the battery
may become depleted while
scanning.
Replace with fully-charged
batteries.
CAN NOT FIND TARGET
PLEASE START OVER
The target center cannot be
found.
The target may not be correctly
collimated or may have been
blocked during scanning.
Reconfirm the position and scan
again. If the error persists, restart
the instrument and scan once
again. If correction is still not
seen, adjustment will be
required.
MEASUREMENT FAILED
PLEASE START OVER
An error occurred during target
center measurement.