TH2826 Operation manual
Ver1.0
◇
35
Operational step of setting high and low limit
1)
Set
Param
,
Nomi
and
Mode.
2)
Move cursor to
Tool
.
3)
Move cursor to
Low limit
of Bin 1. if you select tolerance mode, execute step 4 to step
7
;
if you select sequentional mode, execute step8 to 12.
4)
Use numeric key to input low limit value of Bin 1
,
after inputting data
,
use
(
p
,
n
,
µ
,
m
,
k
,
M
,
*1
)
to replace
[ENTER]
to input limit value. When use
[ENTER]
to input limit
value
,
the default unit of limit value is the same as that of last input. When pressing
*1
,
F
,
H or
Ω
is the default unit. After inputting limit value of bin 1
,
the low limit of bin 1is
set as -(absolute limit )
,
high limit of bin 1is +( absolute limit).
5)
Cursor is jumped to
Low limit
of bin 2
,
repeat step 4 to limit value of bin 9. Then cursor
is jumped to
Low limit
of bin 2.
6)
After inputting low limit value of secondary parameter
,
cursor is jumped to
High limit
of
bin 2
7)
Input the high limit value of secondary parameter.
8)
Use numeric key to input low limit value of Bin 1
,
after inputting data
,
use
(
p
,
n
,
µ
,
m
,
k
,
M
,
*1
)
to replace
[ENTER]
to input limit value. When use
[ENTER]
to input limit
value
,
the default unit of limit value is the same as that of last input. When pressing
*1
,
F
,
H or
Ω
is the default unit.
9)
After inputting low limit value of bin 1
,
cursor is jumped to
High limit
of bin 1.
10)
Cursor is jumped to
High limit
of bin 2.
11)
Repeat step 10 till the input of bin 9, then cursor is jumped to
Low limit
of 2
nd
. Input the
low limit of secondary parameter.
12)
Cursor is jumped to
High limit
of 2
nd
.Input the high limit of secondary parameter.
3.6.7
File management
Refer to 4.2<File management>.
3.6.8
Auxdiary tool
Execute the following step to clear the limit parameter on <Limit list setting> page.
1)
Move cursor to
Tool
.
Clear table
2)
Press
Clear table
,
clear all bin limit value.
3.7
<List sweep setting> page
Press
[LCRZ]
,then
More1/3
,
and
List setting
to enter
<List sweep setting>
page.
As figure 3-11
: