![Tonghui Electronics TH2826/A Operation Manual Download Page 34](http://html.mh-extra.com/html/tonghui-electronics/th2826-a/th2826-a_operation-manual_1126460034.webp)
TH2826 Operation manual
Ver1.0
◇
26
Δ
ABS
=X-Y
Where
,
X: The test value of current DUT.
Y: Preset reference value.
Δ
% mode
(
percentage deviation mode
)
The deviation is the percentage one: the deference of test value and preset reference
value divides the reference value. The formula of calculating
Δ
% is as below
:
Δ
%
=
(
X-Y)
/Y ×100 [%]
Where
,
X: The test value of current DUT.
Y: Preset reference value.
Operational step of deviation test function
Set deviation test according to the operation below.
1)
Move cursor to
Reference A
zone to input the reference value of primary parameter. The
following softkeys are displayed.
Measure
When the component is connected to the test terminal
,
press
Measure
. TH2826 tests
the component, and the result is output automatically to be the value of reference A
and
B
.
2)
Use softkey
Measure
or numeric key to input the reference value of primary parameter.
3)
Move cursor to
(
Reference
)B
zone to input the reference value of secondary
parameter. The following softkeys are displayed.
Measure
When the component is connected to test terminal
,
press
Measure
. TH2826 can test
reference component
,
the result is input as reference A and
B
.
4)
Use softkey
Measure
or numberic key to input the reference value of secondary
parameter. If the primary and secondary parameter have been set on step 2
,
skip this step.
5)
Move cursor to
Deviation A
zone. The following softkeys are displayed.
Δ
ABS
Δ
%
OFF
6)
Use the softkey above, set the deviation mode of primary parameter.
7)
Move cursor to
(
Deviation
)
B
zone. The following softkeys are displayed.
Δ
ABS
Δ
%
OFF
8)
Use the softkey above to set the deviation mode of secondary parameter.
3.4.15
File management
Refer to 4.2<File management>.