TH2826 Operation manual
Ver1.0
◇
114
<value> : NR1,NR2 or NR3 format with HZ, KHZ and MHZ as the shortfix.
<n>:
1:frequency point 1
2:frequency point 2
3:frequency point 3
e.g.
:
WrtCmd(“CORR:SPOT1:FREQ 2KHZ”) ;set frequency 1 to 2 kHz
note
:
the range of <value> must be 20Hz to 200kHz, or the return value will be “error”.
query syntax
:
CORRection:SPOT<n>:FREQuency?
return
:
<NR3><NL^END>
:SPOT<n>:OPEN
is used to perform the open correction for the particular frequency
points(point 1,point 2 and point 3).
command syntax
:
CORRection:SPOT<n>:OPEN
<n>:
1:frequency point 1
2:frequency point 2
3:frequency point 3
e.g.
:
WrtCmd(“CORR:SPOT1:OPEN”) ;perform the open correction for frequency
point 1.
:SPOT<n>:SHORt
is used to perform the short correction for the particular frequency
points(point1, point 2 and point 3).
command syntax
:
CORRection:SPOT<n>:SHORt
<n>:
1:frequency point 1
2:frequency point 2
3:frequency point 3
e.g.
:
WrtCmd(“CORR:SPOT1:SHOR”) ;perform the short correction for point 1
:SPOT<n>:LOAD:STANdard
is used to perform the load correction for the particular
frequency points(point 1, point 2 and point 3) under the standard
reference value.
command syntax
:
CORRection:SPOT<n>:LOAD:STANdard <
REF.A
>,<
REF.B
>
<n>: