TH2826 Operation manual
Ver1.0
◇
32
As figure 3-9
:
Figure 3-9
TH2826 can set bin limit of 9 primary parameters and bin limit value of a secondary parameter.
The test result can be sorted into 10 bins at most
(
BIN1 to BIN9 and BIN OUT
)
. If the
primary parameter is within BIN1 to BIN9
,
but its secondary is not
,
now the DUT is sorted to
Aux bin. When TH2826 installs HANDLER interface
,
applying auto test sorting system, the
compare function is available. The limit parameter can only be set on
<Limit list setting>
page.
(
Parameter
)
(
Mode
)
(
Nominal
)
(
Aux
)
(
Compare
)
(
Low limit
)
(
High limit
)
3.6.1
Swap parameter
This function can swap the primary parameter and secondary parameter. When the parameter
is
:
Cp-D, the parameter swap function can change the parameter as
:
D-Cp, now D can set 9
pairs of compare limit but Cp can set 1 pair of compare limit.
Operational step of swap parameter function
Execute the following steps to swap primary and secondary parameter.
1)
Move cursor to
Parameter
,
the following softkeys are displayed.
Swap parameter
Press
Swap parameter
to swap primary and secondary parameter.
2)
Then press
Swap parameter
to swap primary and secondary parameter to return to the