Note that the power-off state, is considered an abnormal situation.
The battery power is sufficient for another 15 minutes.
As this alarm of lights and sound emits, the user has to:
■
■
Leave the hazardous environment
■
■
Charge the internal battery of the blower unit
■
■
As necessary, charge the external battery
2.3.4 Filter warning or leak
The indicator light that is closest to the on / off button is on.
The other two (2) indicator lights flash.
The blower emits an alarm sound every one (1) second.
This indicates that either:
■
■
The inhalation filter is blocked
■
■
A major leak occurred
In either situation, the user has to:
■
■
Leave the hazardous environment
■
■
Evaluate what caused either the leak or the situation with the filter
■
■
If a situation with the inhalation filter occurred, replace the filter as to:
5.2 Replace the inhalation filter on page 36.
■
■
If a leak occurred due to a mechanical issue, please replace the faulty parts
according to the individual instructions in chapter 3 from page 23.
■
■
If a leak occurred to an unknown reason, refer to:
1.3 Before use on page 5
■
■
If either of the above is the reason for the alarm, contact Tiki Safety AB