ENGLISH – Tiki medical
31
4.4 Calibrate the blower
1. Perform this procedure before using the mask,
hence before entering the environment with the hazardous mixtures
2. To calibrate the blower, push the ‘button for the lights’ (11c) followed by the ‘on /
off button’ (11a on figure 11) for six (6) seconds until the lights stop flashing.
3. If calibrated, the following occurs:
› The red indicator lights flash, refer to 11d on figure 11
4. Test the seal of the mask as to the procedure: 4.6 Testing the mask during use on
page 32
4.5 Adjusting the blower unit speed
1. Press and hold the on/off button for
three seconds to start the blower
unit. The blower unit will then start in
an automatic speed regulated by the
breathing.
2. Press the on/off button one additional
time to change the speed from
automatic speed to low speed.
3. Press the on/off button one additional
time to change the speed from low
speed to medium speed.
4. Press the on/off button one additional
time to change the speed from
medium speed to high speed.
5. Press the on/off button one additional
time to return from high speed to
automatic speed.