38
Tiki medical – ENGLISH
5.4 Storage
The Tiki medical mask can be stored for the following periods to the following
conditions.
These conditions ensure the integrity of the batteries of both the internal battery
and the external battery.
■
■
Long term storage for over (>) six (6) months at 17° C to 28° C at a Relative
Humidity (RH) of under (<) 60%.
Note – The Tiki Blower & External Battery contain LiPo cells and like all LiPO batteries
they must be maintenance charged a minimum once every six (6) months.
Warning – Always supervise all battery charging and stop immediately should the
battery swell, get hot or if it produces smoke or flames.
■
■
Medium term storage for one (1) to six (6) months at 0° C to 38° C at a RH of
under (<) 60%
■
■
Short term storage for less than (<) one (1) month at a temperature between
minus (-) 20° C to 40° C at a RH of less than (<) 30%