ENGLISH – Tiki medical
19
2.4.2 Operational conditions
The Tiki medical mask is designed for use in an operational temperature between
minus (-) 10° C to 45° C and a Relative Humidity (RH) of under 90%.
Do not use the Tiki medical mask in an environment with a RH of over 90%.
Charge the batteries in an environment with a temperature between 5° C to 30°C,
in accordance to: 4.1 Charge the batter(y)(ies) on page 28.
2.4.3 Performance tests
Every Tiki medical mask has been subjected to verification and tests in its production
phase.
As the seal collar is installed on the visor of the mask, the combination is
positioned on an artificial test ‘face’.
In this test configuration a vacuum is applied to the enclosed air volume between
the Tiki medical mask and the face of the user for a pre-set time.
The test measures whether the mask can hold the vacuum for the pre-set time.
Only masks that hold the vacuum for the pre-set time are validated and further
processed in the production.
Masks that do not hold the vacuum for the pre-set time are not validated and not
further processed in the production.
The test installation generates information on the tests and saves this information.
2.4.4 Traceability
As every Tiki medical mask holds a unique number, the information of the respective
performance test of each mask is traceable from the production to the end-user.
Please mark that this information is only available for legal incident investigation
on request to Tiki Safety AB.
Also refer to the procedure to read the Quick Response (QR) code at:
3.3 Read the QR code on page 26.