Portable Optical Tweezers Kit
Chapter 5: Setup and Adjustment
Rev B, February 15, 2022
Page 37
Camera System Components
Slip lens 3 from Figure 23 onto the cage rods and secure it with the locking screws on the
side. Make sure the locking screws do not point towards the side where the damped post
will be (i.e., the side of the C1498(/M) clamp).
Next, slip on the filter in the cage plate from Figure 24. Also make sure no locking screws
face towards the C1498(/M) clamp. The cage rods should be inserted into the CP33T(/M)
only half way
. Make sure that the upper locking screws fix the cage rods to the thick cage
plate. The lower locking screws
need to be left free (so the camera
segment can be attached to the
beamsplitter assembly). The fully
assembly camera segment can be
seen in Figure 26.
The SM1NT1 slotted locking ring
coupler allows the user to choose
the optimal camera orientation.
Hold the camera in position with
one hand and use the SPW502
spanner wrench to fix the camera
such that the camera cable is on
the left side. If the camera is not
parallel to the cage plates, the
motion of the stage will result in a
skewed motion of the camera
image.
Double check that the orientation of
all cage plates and the camera is
identical to Figure 26.
Camera System
Summary of Contents for EDU-OT3
Page 1: ...EDU OT3 EDU OT3 M Portable Optical Tweezers User Guide...
Page 93: ......