BC207 Series
Chapter 6: Operation Instructions
Rev: 8.1, 10-Jan-2022
Page 76
© 2022 Thorlabs
6.2.2 Pass/Fail Test
The Calculation Results panel includes a pass/fail test for reproducible assessment of light
beams.
For each parameter a minimum and / or maximum can be set as criteria.
Pass / fail test criteria can be set to "not below minimum", "not above maximum" by setting the
check at Min. or Max., or "between minimum and maximum" by checking the appropriate boxes
both, the
Min.
and
Max.
columns.
Test results will be displayed in the column
Test
only for those parameters that are selected as
test criteria:
The selected test criteria was fulfilled; test passed.
The test criteria
Min.
was underrun; test failed
The test criteria
Max.
was exceeded; test failed
Note
As per definition, the beam ellipse has a major and a minor axis. For a pass / fail test, a mini-
mum and maximum value can be entered in the fields for minor and major ellipse diameter. In
the example given above, the "pass" ranges are:
- Minor axis diameter must be between
Min.
= 2600 and
Max.
= 2800 µm
- Major axis diameter must be between
Min.
= 2600 and
Max.
= 2800 µm
The test is passed only if both conditions are fulfilled.
Summary of Contents for BC207UV
Page 2: ...Copyright 2022 Thorlabs Version Date 8 1 10 Jan 2022 ...
Page 103: ...BC207 Series Chapter 6 Operation Instructions Rev 8 1 10 Jan 2022 Page 96 2022 Thorlabs ...
Page 133: ...BC207 Series Chapter 6 Operation Instructions Rev 8 1 10 Jan 2022 Page 126 2022 Thorlabs ...
Page 181: ...www thorlabs com ...