BC207 Series
Chapter 6: Operation Instructions
Rev: 8.1, 10-Jan-2022
Page 100
© 2022 Thorlabs
·
Align the beam position using the controls of your laser source. Subsequently click the cap-
ture button of Position 1 and 2, observe the alignment, and improve it until all four alignment
criteria are fulfilled (bulbs must be at least yellow):
Well aligned beam position
·
To toggle between the two positions, the movement of the stage between the two stop po-
sitions can be controlled automatically by the software. Just push the
loop button
and
enter a convenient dwell time value. The stage starts a loop move, dwelling at the stop po-
sition for the given time. During the dwelling a realignment can be made. Please keep in
mind, that the numeric alignment indicators are being updated only after the next move.
Summary of Contents for BC207UV
Page 2: ...Copyright 2022 Thorlabs Version Date 8 1 10 Jan 2022 ...
Page 103: ...BC207 Series Chapter 6 Operation Instructions Rev 8 1 10 Jan 2022 Page 96 2022 Thorlabs ...
Page 133: ...BC207 Series Chapter 6 Operation Instructions Rev 8 1 10 Jan 2022 Page 126 2022 Thorlabs ...
Page 181: ...www thorlabs com ...