![THORLABS BC207UV Operating Manual Download Page 144](http://html.mh-extra.com/html/thorlabs/bc207uv/bc207uv_operating-manual_1110819144.webp)
BC207 Series
Chapter 6: Operation Instructions
Page 137
MTN026962-D02
While running the measurement, most of the buttons and options are disabled, e.g. the Diver-
gence measurement settings and the toolbar. This prevents the modification of settings during
a measurement.
If necessary, the measurement can interrupted by clicking the
Stop
button
.
After starting the measurement the X axis of the graph is adapted to the user-defined scan
range, for example from 0 to 120 mm. The Y axis scales automatically to the recorded beam
widths.
6.3.8.6 Divergence Measurement Results
The green bulb indicates that the measurement was successful.
These values can also be found in the listing of the complete
Results
.
Negative values indicate a converging beam, while positive values indicate a diverging beam.
Note
All results are calculated from the applied linear fit!
Full Divergence Angle X' (Y')
is explained in chapter
.
Divergence Asymmetry
Asymmetry is the quotient of divergence angles in Y and X scan direc-
tions. Values differing from 1.0 indicate that the beam ellipticity is changing with z position, for
instance an elliptical beam is focused to a round spot.
140
Summary of Contents for BC207UV
Page 2: ...Copyright 2022 Thorlabs Version Date 8 1 10 Jan 2022 ...
Page 103: ...BC207 Series Chapter 6 Operation Instructions Rev 8 1 10 Jan 2022 Page 96 2022 Thorlabs ...
Page 133: ...BC207 Series Chapter 6 Operation Instructions Rev 8 1 10 Jan 2022 Page 126 2022 Thorlabs ...
Page 181: ...www thorlabs com ...