OQ/PQ Operating Instructions
M1393R0
Page 57 of 60
6.8
Switchos Pump Test (NAN Configuration)
Sequence: XQ_7_Switchos Pump Test_NAN
Program:
;Program file for switchos pump stability test
;Program version 08/10/2004
;Settings for restrictor (320cm, 15um ID)
Diameter
=
75um
Length
=
15cm
StationaryPhase
=
C18_3um_100A
;CRP
=
625
CalibrateCRP
When=BeforeFirstSample
;ParkPercentage
=
Disabled
;Pump limits settings
MasterPressure.LowerLimit
= 0.0
MasterPressure.UpperLimit
= 400.0
ColumnPressure.LowerLimit
= 0.0
ColumnPressure.UpperLimit
= 200.0
TrapColumnPressure.LowerLimit
=
0.0
TrapColumnPressure.UpperLimit
=
200.0
;Settings for UV detector
Data_Collection_Rate
=
2
UV_VIS_1.Wavelength =
254
TimeConstant
=
2
;FAMOS
settings
InjectMode
=
Fullloop
LowDispersionMode
=
Off
UseAirSegment
=
Off
UseHeadSpace
=
Off
RinseBetweenReinjections
= Yes
SyringeSpeed
=
Low
SyringeSpeedFactor
=
0.1
SampleHeight
=
4
FlushVolume
=
5.0
WashVolume
=
50
;Commands for FAMOS with cooling option
;Sampler.TempCtrl =
On
;Sampler.Temperature.Nominal =
20.00
;Sampler.Temperature.LowerLimit = 5.00
;Sampler.Temperature.UpperLimit =
30.00
;Virtual channel settings
pump_pressure.Formula
Formula=masterpressure.value
pump_pressure.Type
=
Analog
column_pressure.Formula
Formula=pump.columnpressure.value
column_pressure.Type
=
Analog
Loading_Pump.TrapColumnPressure.LowerLimit
=
0.0
Loading_Pump.TrapColumnPressure.UpperLimit
=
400.0
Trap_Column_Pressure.Formula
Formula=Loading_Pump.TrapColumnPressure.value
Trap_Column_Pressure.Type =
Analog