CHROMELEON
Listings
OQ/PQ Operating Instructions
M1393R0
Page 51 of 60
6.5
Autosampler Reproducibility (NAN Configuration)
Sequence: XQ_4_Autosampler Reproducibility_NAN
Wait
Sampler.Ready
;Program file for injection reproducibility test
;System must be setup for Nano LC experiments
;For this test time constant is 0.5 instead of 2.0
;Program version 08/10/2004
;Settings for restrictor (320cm, 15um ID)
Diameter
=
75um
Length =
15cm
StationaryPhase =
C18_3um_100A
;CRP
=
625
CalibrateCRP
When=BeforeFirstSample
;ParkPercentage =
Disabled
;Pump limits settings
MasterPressure.LowerLimit =
0.0
MasterPressure.UpperLimit =
400.0
Pump.columnPressure.LowerLimit = 0.0
Pump.columnPressure.UpperLimit = 200.0
TrapColumnPressure.LowerLimit
=
0.0
TrapColumnPressure.UpperLimit
=
200.0
;Settings for UV detector
Data_Collection_Rate =
2
UV_VIS_1.Wavelength =
272
TimeConstant
=
0.5
;FAMOS
settings
Wait
Sampler.Ready
InjectMode
=
Partial
LowDispersionMode =
Off
UseAirSegment =
Off
UseHeadSpace =
Off
RinseBetweenReinjections
= Yes
SyringeSpeed
=
Low
SyringeSpeedFactor =
0.1
SampleHeight =
9
FlushVolume
=
5.0
WashVolume =
50
;Commands for FAMOS with cooling option
;Sampler.TempCtrl =
On
;Sampler.Temperature.Nominal =
20.00
;Sampler.Temperature.LowerLimit = 5.00
;Sampler.Temperature.UpperLimit =
30.00
;Virtual channel settings
pump_pressure.Formula
Formula=masterpressure.value
pump_pressure.Type
=
Analog
column_pressure.Formula
Formula=pump.columnpressure.value
column_pressure.Type
=
Analog
;Ultimate pump settings
Pump.%A.Equate =
"100% water + 0.1% FA"