Changing Ionization Modes
Removing the Ion Volume
Thermo Scientific
TSQ Quantum XLS Series User Guide
79
9. Remove the ion volume from the I/R tool (see
a. Wearing clean gloves, press the ion volume forward into the tip of the I/R tool and
rotate it to disconnect the bayonet pins from the pin guides.
b. Pull the ion volume out of the I/R tool.
Figure 48.
Ion volume, ion volume holder, and I/R tool
10. To clean the ion volume, use the instructions in
“Cleaning Stainless Steel Parts”
. If you are cleaning a CI ion volume, be sure to clean out the small electron
entrance hole. Aluminum oxide can get trapped in this hole, which can adversely affect
sensitivity. Use a dental pick or old syringe needle to clean the hole.
WARNING
Burn Hazard
. Because the ion volume might be too hot to touch, let
it cool to room temperature before handling it.
Arrow
Bayonet
pin
Bayonet
lock
Arrow
I/R tool
Bayonet pin
guide
Ion volume holder
Spring washer
Ion volume