5
Changing Ionization Modes
Removing the Ion Volume
76
TSQ Quantum XLS Series User Guide
Thermo Scientific
Figure 44.
I/R tool at the first stop on guide bar and the inlet valve knob tightened
6. Click
OK
. The forepump evacuates the inlet valve. Wait until you can safely insert the
probe (
), and then click
OK
.
Figure 45.
Safe to insert the probe message
7. Once evacuation is complete, pull the inlet valve lever up (
Inlet valve knob
First stop of
guide bar
I/R tool