Functional Description
Mass Spectrometer
22
TSQ Quantum XLS Series User Guide
Thermo Scientific
•
•
•
Vacuum System and Inlet Gasses Hardware
•
•
Controls and
Indicators
Five light-emitting diodes (LEDs) are located in the upper right of the front panel of the mass
spectrometer. See
The Power LED illuminates green whenever power is supplied to the vacuum system and
electronic assemblies of the mass spectrometer.
The Vacuum LED illuminates yellow when the turbomolecular pump is nearly at speed
(80 percent of its operating speed of 750 MHz), and it is safe to turn on the ion gauge. The
Vacuum LED is off if the turbomolecular pump is not at speed. The Vacuum LED
illuminates green whenever the pressure in the analyzer chamber, as measured by the ion
gauge, is at or below the value required to enable high voltages to the mass analyzer.
See
Figure 10.
Front panel LEDs of the TSQ Quantum XLS Ultra mass spectrometer
Figure 11.
Front panel LEDs of the TSQ Quantum XLS mass spectrometer
Table 2.
Maximum allowed pressure to turn on high voltages
Carrier gas
Ar collision gas
Maximum pressure (Torr)
He
Off
8 × 10
–
6
He On
5 × 10
–
5
Power
Vacuum
Communication
System
Scan
Communication
Vacuum
Power
System
Scan