6
Maintenance
Cleaning Ion Source Components
Thermo Scientific
TSQ Quantum XLS Series User Guide
87
Cleaning Ion Volumes
You do not need to vent the mass spectrometer to atmosphere to clean the ion volume.
Y
To clean the ion volume
1. Remove the ion volume. See
.
2. Clean the ion volume. See
“Cleaning Stainless Steel Parts”
. If you are cleaning
a CI ion volume, be sure to clean out the small electron entrance hole. Aluminum oxide
can get trapped in this hole, which can adversely affect sensitivity. Use a dental pick or old
syringe needle to clean the hole.
3. Reinstall the ion volume. See
.
Cleaning the Lenses of the Ion Source
If cleaning the ion volume does not restore system performance, clean the ion source lenses.
Y
To clean the ion source lenses
1. Shut down and vent the system. See
“Shutting Down the System Completely”
2. Remove the ion source. See
3. Remove the ion source lens assemblies from the ion source and disassemble it. See
“Removing the L1, L2, L3 Lens Assembly and the Heated Q0 Quadrupole Assembly”
4. Clean the lenses and the spacers. See
“Cleaning Stainless Steel Parts”
and
“Cleaning Non-Stainless Steel or Hybrid Parts”
, respectively.
5. Reassemble L1, L2, L3 lens assembly and install it and the heated Q0 quadrupole onto
the ion source. See
“Reassembling the L1, L2, L3 Lens Assembly”
WARNING
Burn Hazard
. The ion volume might be hot. Allow the ion volume to
cool to room temperature before touching it.
WARNING
Shock Hazard
. Unplug the TSQ Quantum XLS Series system before
proceeding.
WARNING
Burn Hazard
. The ion source might be hot. Allow the ion source to cool
to room temperature before touching it.