Maintenance
Cleaning Ion Source Components
Thermo Scientific
TSQ Quantum XLS Series User Guide
91
Disassembling the L1, L2, L3 Lens Assembly
Y
To disassemble the L1, L2, L3 lens assembly
1. Remove the lens retainer clip by pinching the ends with your fingers or a pair of tweezers.
See
2. Remove the lenses and spacers from the lens holder.
3. Place the components on a clean surface.
Figure 53.
L1, L2, L3 lens assembly, assembled and exploded views
Lens holder
Lens L3
Lens spacer
Lens L2
Lens spacer
Lens L1
Lens retainer clip
Note
To clean the ion source lenses, follow the procedure in
. To clean the remaining parts, follow the procedure in