Step
Touchscreen operation
Screen displays and notes
4
To set the start time, press
the
[Start Time]
button and to
set the stop time, press the
[Stop Time]
button. Enter the
start time and stop time from
the on-screen keypad.
Pressing the
[Start Time]
button will display the
[Start
Time]
in the set state display and pressing the
[Stop
Time]
will display the
[Stop Time]
in the set state display.
Pressing the input fields will make them ready to accept
entries.
Use the
[<]
and
[>]
buttons to switch between items
(month, day, hour, minute).
Enter a numeric between 0 and 23 (24-hour clock) for
[Hour].
Enter a date and time later than the current time. The stop
time should take centrifugation time into account and
centrifugation start time should be set to a time later than
current time.
5
Press the
[Enter]
button.
6
Press the
[START]
button.
The vacuum pump goes on and temperature control starts.
The RTC display screen shows the START time and STOP
time indicating that RTC operation is ready to start.
The centrifuge will start automatically at the set time.
Set state display
●
A message
prompting you to
check operation
appears.
RTC will not start unless
you press the
[START]
button.
The
[STOP]
button allows
you to cancel the RTC
operation setting.
3-3-4 RTC (Real Time Control) Operation
(Upper row)
Operation start time
(Lower row)
Operation stop time
RTC display
3-56