Step
Touchscreen operation
Screen displays and notes
9
Press the field in Step 1,
check that the field turns
blue and press the
[Edit]
button.
10 Press the white area in the
SPEED setting value display
field ((1)).
(You can press the TIME
((2)) or TEMP ((3)) setting
value display fields and
make similar settings.)
11 Enter SPEED, TIME and
TEMP run parameters.
Press the display field of the item you want to set, enter run
parameters using the on-screen keypad and press the
[Enter]
button.
The
[Enter]
button is replaced by the
[OK]
button.
For details, refer to Section 3-2-1 "Setting Run Parameters."
Example)
SPEED: 100,000 rpm
TIME: 10:00
TEMP
:
10.0
°
C
●
The
run parameter
entry/display screen
appears.
(1) SPEED setting value
display field
(2) TIME setting value
display field
(3) TEMP setting value
display field
(4) ACCEL setting value
display field
(5) DECEL setting value
display field
(6) RCFmax setting value
display field
(7) RCFavg setting value
display field
(8) ω
2
T setting value
display field
●
SPEED, TIME, TEMP
display fields and the
on-screen keypad
appear.
NOTE
1) A SPEED entry is not
required when an RCF
value is set.
2) A TIME entry is not
required
when a ω
2
T
value is set.
[Edit]
button
3-3-1-1 Program Registration Procedures
(1)
(2)
(3)
(4)
(5)
(6)
(7
)
(8)
3-27