Step
Touchscreen operation
Screen displays and notes
3
Press the RCF display screen
field.
The RCF setting screen displays the RCFmax and
RCFavg values computed from the currently displayed
SPEED value and the rotor set in step 1.
4
Press the field of the item you
want to set.
In this example, 475,000 x g
will be set as the RCFmax
value that is displayed when
the RCF field is pressed.
5
Press the on-screen keypad to
enter a setting value.
All except the last
two digits turn blue.
●
The set value turns
blue indicating it is
waiting for user
input.
Calculated RCFavg
value
●
The RCF setting
screen appears.
Set rotor model
Calculated
RCFmax value
●
The entered value is
displayed in blue.
On-screen keypad
[Enter]
button
3-3-2 RCF Display and Setting Function