3-11
Graphical User Interface
3. 3. Pre-Temper the Centrifugation Chamber
Refrigerated centrifuges allow for pre-tempering, that is pre-warming or pre-cooling, the centrifugation chamber
and the empty rotor before the centrifugation run starts. If necessary pre-temper your samples using proper
equipment. The centrifuge is not intended to be used to pre-temper your samples.
NOTICE
Ventilated models cannot pre-temper the centrifugation chamber.
Proceed as follows to set the pre-tempering target temperature for the centrifuge:
1. Tap the
temperature box
on the “Home” screen, shown in
Figure 3–22: Temperature Box on Home Screen
The “Setpoints: Standard” screen shown in
or “Setpoints: Advanced” screen shown in
appears.
2. Tap in the
Temp
entry field, then use the keypad on the right to enter the desired target temperature.
The admissible temperature range for the unit appears just below the “Temp” label to assist you in setting
target temperature correctly.
When you tap the
Temp
entry field, the previous setting will be replaced immediately as you start entering
digits on the keypad.
3. Tap the
Save
button to save the temperature setting for the pre-tempering run.
You will be returned to the “Home” screen. The new target temperature appears as the setpoint below the
current temperature.
4. Tap the
Pre-Temp
button on the “Home” screen and then the
Start
button to start pre-tempering.
The centrifuge starts heating or cooling the centrifugation chamber to the setpoint temperature.
The current centrifugation chamber temperature displayed on top of the setpoint temperature starts to
change towards the setpoint.
5. When the centrifugation chamber temperature reaches the pre-tempering setpoint, the “Pre-Temp Complete”
pop-up window shown in
appears, announcing that the centrifuge is correctly pre-tempered
and ready.
Figure 3–23: Pre-Temp Complete Pop-up Window
6. Tap the
Pre-Temp
button to exit the Pre-Temp mode.
Summary of Contents for Sorvall ST Plus Series
Page 116: ...B 9 TX 750 B 1 4 Biocontainment Certificate ...
Page 120: ...B 13 TX 1000 B 2 4 Biocontainment Certificate ...
Page 126: ...B 19 HIGHConic II B 4 4 Biocontainment Certificate ...
Page 133: ...B 26 Fiberlite F15 6 x 100y B 7 4 Biocontainment Certificate ...
Page 137: ...B 30 HIGHPlate 6000 B 9 4 Biocontainment Certificate ...
Page 140: ...B 33 M 20 Microplate B 10 4 Biocontainment Certificate ...
Page 143: ...B 36 Microliter 48 x 2 B 11 4 Biocontainment Certificate ...
Page 146: ...B 39 Microliter 30 x 2 B 12 4 Biocontainment Certificate ...
Page 149: ...B 42 MicroClick 30 x 2 B 13 4 Biocontainment Certificate ...
Page 152: ...B 45 MicroClick 18 x 5 B 14 4 Biocontainment Certificate ...
Page 155: ...B 48 Fiberlite F21 48 x 2 B 15 4 Biocontainment Certificate ...
Page 159: ...B 52 Fiberlite F10 6 x 100 LEX B 17 4 Biocontainment Certificate ...