17
PerpHecT
®
Orion 310, 320, 330 Instruction Manual
B. Orion 310 Temperature Calibration Procedure
This procedure should be performed at least once per week to ensure
accurate temperature compensation when operating in the LogR mode.
If the meter is powered down or the electrode disconnected, it is
recommended that you recalibrate the LogR temperature. See the
Appendix Section for accuracy specifications and recommended
calibration techniques.
Choose a temperature calibration range which is at least 5
°
C apart
depending on your expected operating range. Evenly bracket your
expected sample temperature range. A maximum 20
°
C range is
recommended for maximum accuracy. For example, for measurement
around 20
°
C calibrate at 10
°
C and 30
°
C. For temperature span greater
than 20
°
C, a three-point temperature calibration is recommended.
For many samples, a single-point calibration near the expected sample
temperature will yield adequate results. A constant temperature bath or
a hot plate is recommended to hold the temperature of calibrating solution.
For best results during temperature calibration, it is recommended
that pH buffer or tap water be used as the temperature calibrating solution.
Do Not Use
deionized water because the conductivity is too low to obtain a
good calibration.
IMPORTANT:
When using the meter for the first
time or restarting in the LogR mode, you must
perform a LogR temperature calibration. It is
recommended that at least a two-point temperature
calibration be performed to meet optimum system
accuracy. See Appendix Section for additional
information.
Introduction to Patented digital LogR pH Measurement for Orion 310
Summary of Contents for Orion 310
Page 12: ...PerpHecT Orion 310 320 330 Instruction Manual 8 ...
Page 18: ...PerpHecT Orion 310 320 330 Instruction Manual 14 ...
Page 24: ...PerpHecT Orion 310 320 330 Instruction Manual 20 ...
Page 26: ...PerpHecT Orion 310 320 330 Instruction Manual 22 ...
Page 30: ...PerpHecT Orion 310 320 330 Instruction Manual 26 ...
Page 40: ...PerpHecT Orion 310 320 330 Instruction Manual 36 ...
Page 42: ...PerpHecT Orion 310 320 330 Instruction Manual 38 ...
Page 46: ...PerpHecT Orion 310 320 330 Instruction Manual 42 ...
Page 56: ...PerpHecT Orion 310 320 330 Instruction Manual 52 ...
Page 64: ...PerpHecT Orion 310 320 330 Instruction Manual 60 Troubleshooting Guide ...
Page 66: ...PerpHecT Orion 310 320 330 Instruction Manual 62 ...
Page 76: ...PerpHecT Orion 310 320 330 Instruction Manual 72 ...
Page 78: ...PerpHecT Orion 310 320 330 Instruction Manual 74 ...