The Operator Assistance Codes E-20 through E-36 alert the user to a
potential problem while calibrating or measuring. Several steps can be
taken to eliminate the problem in each case. If the code persists after
trying the suggested remedies, contact The Technical Edge
SM
for Orion
Products at 1-800-225-1480.
NOTE:
Press any key to clear an Error message
from the display
Code
Description
Causes / Suggestions
E-20
Out of Range
1. Occurs when electrode is out of
solution. Code will disappear when
electrode is returned to solution.
2. Sample may be out of range.
Check system using a buffer.
3. Recalibrate system using fresh buffers.
4. Refer to electrode instruction manual
for instructions on how to clean the
electrode.
E-21
ABR error/
Electrode voltage being measured is
pH AutoCal error
>
±
0.5 pH units from the nominal value
for the pH buffer. Press key to
acknowledge clear error.
1. Verify buffer is pH 4.01, 7.00 or 10.01.
2. Recalibrate using fresh buffers.
3. Refer to pH electrode instruction
manuals for instructions on how to
clean the electrode.
4. Verify that the buffer the meter is
prompting for is being measured.
5. Perform manual calibration.
NOTE:
Non-standard calibration buffers
may be used only for manual calibrations.
Operator Assistance Codes Orion 310, 320 and 330
Summary of Contents for Orion 310
Page 12: ...PerpHecT Orion 310 320 330 Instruction Manual 8 ...
Page 18: ...PerpHecT Orion 310 320 330 Instruction Manual 14 ...
Page 24: ...PerpHecT Orion 310 320 330 Instruction Manual 20 ...
Page 26: ...PerpHecT Orion 310 320 330 Instruction Manual 22 ...
Page 30: ...PerpHecT Orion 310 320 330 Instruction Manual 26 ...
Page 40: ...PerpHecT Orion 310 320 330 Instruction Manual 36 ...
Page 42: ...PerpHecT Orion 310 320 330 Instruction Manual 38 ...
Page 46: ...PerpHecT Orion 310 320 330 Instruction Manual 42 ...
Page 56: ...PerpHecT Orion 310 320 330 Instruction Manual 52 ...
Page 64: ...PerpHecT Orion 310 320 330 Instruction Manual 60 Troubleshooting Guide ...
Page 66: ...PerpHecT Orion 310 320 330 Instruction Manual 62 ...
Page 76: ...PerpHecT Orion 310 320 330 Instruction Manual 72 ...
Page 78: ...PerpHecT Orion 310 320 330 Instruction Manual 74 ...