19
PerpHecT
®
Orion 310, 320, 330 Instruction Manual
Chapter VII
Temperature Measurement
for Orion 310
A. Temperature Compensation
(Manual and Automatic)
1. Use an ATC probe (Orion 927005 or Orion 927006) for automatic
temperature compensation. Disable the LogR function using the
TEMP key.
2. Attach ATC probe to 3.5 mm phone jack on rear panel of the
PerpHecT pH meter. See
Figure 4
.
3. For manual temperature compensation, disconnect the ATC probe.
Manual temperature may only be set when the LogR function is
disabled and there is no ATC probe connected:
Press the MODE key until the temp mode is entered. Use the
SCROLL (
▲
,
▼
) keys to manually set the temperature value. Press
the MODE key to return to pH mode. This temperature will now
be used to calculate temperature corrected pH values.
4. To return to ATC mode, attach the ATC probe to the ATC jack
(see
Figure 4
). Temperature compensation will now be automatic.
Temperature Measurement for Orion 310
Summary of Contents for Orion 310
Page 12: ...PerpHecT Orion 310 320 330 Instruction Manual 8 ...
Page 18: ...PerpHecT Orion 310 320 330 Instruction Manual 14 ...
Page 24: ...PerpHecT Orion 310 320 330 Instruction Manual 20 ...
Page 26: ...PerpHecT Orion 310 320 330 Instruction Manual 22 ...
Page 30: ...PerpHecT Orion 310 320 330 Instruction Manual 26 ...
Page 40: ...PerpHecT Orion 310 320 330 Instruction Manual 36 ...
Page 42: ...PerpHecT Orion 310 320 330 Instruction Manual 38 ...
Page 46: ...PerpHecT Orion 310 320 330 Instruction Manual 42 ...
Page 56: ...PerpHecT Orion 310 320 330 Instruction Manual 52 ...
Page 64: ...PerpHecT Orion 310 320 330 Instruction Manual 60 Troubleshooting Guide ...
Page 66: ...PerpHecT Orion 310 320 330 Instruction Manual 62 ...
Page 76: ...PerpHecT Orion 310 320 330 Instruction Manual 72 ...
Page 78: ...PerpHecT Orion 310 320 330 Instruction Manual 74 ...