Orion 310, 320, 330 Instruction Manual
58 Operator Assistance Codes Orion 310, 320 and 330
Code
Description
Causes / Suggestions
E-35
Bad LogR
Displayed when two calibration standards
Standard
do not differ in temperature by the
minimum allowable differences in
LogR mode.
1. The difference between temperature
points must be at least 5
°
C.
E-36
No LogR Cal
Displayed when the calibration
coefficients from a LogR calibration
are out-of-range. Incorrect temperature
value entered.
1. Recheck temperatures and recalibrate.
2. Check electrode and inputs
E-40
Serial Memory
Displayed at power-up when an error
Error
occurs accessing factory calibration data.
1. Unplug meter, plug in again after
10 seconds.