Installation
Operational check
Thermo Fisher Scientific
Thermo Scientific Multidrop Combi nL User Manual
23
Figure 3–13.
Reagent reservoir connected to the instrument
Complete the following procedure to confirm the correct functioning of
the dispenser prior to normal use.
1.
Connect the instrument to a correctly installed line power outlet
which has a protective conductor also called earth or ground.
Switch the instrument on using the mains switch (Figure 2–4).
Warning
Interruption of the protective conductor inside or outside the
instrument or disconnection of the protective conductor terminal may
make the instrument dangerous.
s
l
If the instrument starts properly:
l
The display lights up.
l
The plate carrier moves to the home position and the
dispensing valve lifting mechanism to the up position.
l
The clamp (Figure 3–14) opens.
Operational
check