Technical Specifications
Remote control to Multidrop Combi nL
88
Thermo Scientific Multidrop Combi nL User Manual
Thermo Fisher Scientific
Note
1) The
installation category
(overvoltage category) defines the level
of transient overvoltage which the instrument is designed to withstand
safely. It depends on the nature of the electricity supply and its
overvoltage protection means. For example, in CAT II which is the
category used for instruments in installations supplied from a supply
comparable to public mains, such as hospital and research laboratories
and most industrial laboratories, the expected transient overvoltage is
2500 V for a 230 V supply and 1500 V for a 120 V supply.
2) The
pollution degree
describes the amount of conductive pollution
present in the operating environment. Pollution degree 2 assumes that
normally only nonconductive pollution, such as dust, occurs with the
exception of occasional conductivity caused by condensation.
s
The Multidrop Combi nL can be used as part of an automated system.
When the Multidrop Combi nL is in remote mode, the main view is as
shown below. For more details, refer to the
Multidrop Combi nL Remote
Control Command Sets
(Cat. no. D05871) document. Contact your
distributor or local Thermo Fisher Scientific representative to obtain
the document in PDF format.
The instrument has two alternative computer interfaces: RS-232 and
USB.
Note
It is recommended to use USB to improve the reliability of traffic
in robot applications.
s
Note
The interface to use is selected from the Multidrop Combi nL
user interface.
s
Remote control to
Multidrop Combi nL
Summary of Contents for Multidrop Combi nL
Page 1: ...Thermo Scientific Multidrop Combi nL User Manual Rev 1 2 Cat no N07171 ...
Page 100: ...Glossary 100 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Page 104: ...104 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Page 105: ...Thermo Fisher Scientific Thermo Scientific Multidrop Combi nL User Manual 105 Notes ...
Page 106: ...106 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific Notes ...