Maintenance
Regular and preventive maintenance
Thermo Fisher Scientific
Thermo Scientific Multidrop Combi nL User Manual
77
Figure 5–30.
Replacing the dispensing valve
6.
Fasten the valve screw.
7.
Reattach the dispensing valve tubing (Figure 5–26).
8.
Finally prime with distilled and filtered laboratory-grade water to
ensure the proper functioning of the dispensing valve.
9.
Check the valve correction factor marked on the inspection sheet
enclosed with the spare valve and use FILLit Software for Multidrop
Combi nL to set the appropriate value in the valve correction menu.
Open FILLit Software and select Instrument >
Valve correction
.
Enter the given value in the respective valve (channel) position and
press
Update
to save the value. See the details of FILLit Software in
the
FILLit Software for Multidrop Combi nL User Manual
(Cat. no.
N17562) provided with the instrument.
Note
Check that your instrument has the embedded software version
1.00.41 or greater to successfully utilize the valve correction factors.
s
Note
The instrument does not have to be recalibrated when you
change a dispensing valve, because the valve performance is
consistent.
s
To replace a tubing set and a manifold (Figure 5–31):
Replacing a tubing set
and a manifold
Summary of Contents for Multidrop Combi nL
Page 1: ...Thermo Scientific Multidrop Combi nL User Manual Rev 1 2 Cat no N07171 ...
Page 100: ...Glossary 100 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Page 104: ...104 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Page 105: ...Thermo Fisher Scientific Thermo Scientific Multidrop Combi nL User Manual 105 Notes ...
Page 106: ...106 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific Notes ...