Routine Operation
Protocol settings
46
Thermo Scientific Multidrop Combi nL User Manual
Thermo Fisher Scientific
Use the
Up
and
Down
arrow keys to select the dispensing height
and/or the dispensing offset.
Press
OK
.
The default dispensing height depends on the plate type
(see Table 4–3). You can change the dispensing height values if the
plate used is not the standard one specified in the plate type list.
Use the
Up
and
Down
arrow keys to set the dispensing height. While
you change the offset value, the dispensing valve lifting mechanism
moves * accordingly. The height is set in 0.05 mm increments in both
the up and down directions. The minimum height value depends on
the selected plate type (see Table 4–3). The default dispensing height is
1 mm above the selected plate.
Note
You can adjust the dispensing height of low profile plates below
the plate height.
s
Press
OK
. If the
OK
button is not pressed within 60 seconds, the
adjustments are disregarded and the main view returns to the
Settings
menu.
* z-axis = the dispensing valve lifting mechanism moves
Height