Routine Operation
Running a protocol
50
Thermo Scientific Multidrop Combi nL User Manual
Thermo Fisher Scientific
Note that multistep protocols, that is, protocols with several steps
(e.g. Dispense, Shake, Empty, etc.), can be run with FILLit Software for
Multidrop Combi nL. You can also download protocols with one
dispense step and an additional shake step to the instrument. For more
information, refer to the
FILLit Software for Multidrop Combi nL User
Manual
(Cat. no. N17562).
To run a dispensing protocol, follow the steps below.
Note
Do not dispense extensively without any liquid to avoid
overheating of the pump.
l
Check that there is always enough liquid in the reagent reservoir
and ensure that the reagent filter is below the liquid level. Use a test
tube placed inside the reagent reservoir if a limited amount of liquid
is available.
l
Check after each dispensing that the priming vessel is not filled up
if the tube drain is not used.
s
1.
Make sure you are in the
Main
menu.
2.
Select the plate type.
3.
Select the dispensing protocol from the
protocol list. Only protocols for the selected
plate type are shown.
4.
Press the
PRESSURE
button and wait till the
pressure has risen.
Only required at startup!
5.
Keep the
PRIME
button down until the
tubings are completely filled.
6.
Press the
START
button to start the protocol.
Running a
protocol
Summary of Contents for Multidrop Combi nL
Page 1: ...Thermo Scientific Multidrop Combi nL User Manual Rev 1 2 Cat no N07171 ...
Page 100: ...Glossary 100 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Page 104: ...104 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Page 105: ...Thermo Fisher Scientific Thermo Scientific Multidrop Combi nL User Manual 105 Notes ...
Page 106: ...106 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific Notes ...