Installation
Operational check
24
Thermo Scientific Multidrop Combi nL User Manual
Thermo Fisher Scientific
Figure 3–14.
Priming vessel inserted into place
2.
Ensure that the priming vessel is inserted correctly into its slot
3.
Pull the protective cover over the dispensing valve head.
4.
Press
PRESSURE
.
l
The pump starts and the instrument raises the pressure.
l
The light indicator blinks while the instrument regulates the
pressure.
l
The light is stable once the pressure has been reached.
Caution
Unfastening the tubing while the pressure is on can cause a
hazardous situation, flood or spillage.
l
Check that there is always enough liquid in the reagent reservoir
and ensure that the filter is below the liquid level.
l
Check after each dispensing that the priming vessel is not filled up
if the tube drain is not used.
s
5.
Press
PRIME
.
l
The lifting mechanism is lowered and the dispensing valves are
opened as long as you press
PRIME
.
Priming vessel –
with a
crest
in the middle where the
priming liquid is aimed at
to avoid splashing and a
drain
Safety grooves –
designed to slope and prevent
fluid from entering the instrument
Clamp
Light indicator
Summary of Contents for Multidrop Combi nL
Page 1: ...Thermo Scientific Multidrop Combi nL User Manual Rev 1 2 Cat no N07171 ...
Page 100: ...Glossary 100 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Page 104: ...104 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific ...
Page 105: ...Thermo Fisher Scientific Thermo Scientific Multidrop Combi nL User Manual 105 Notes ...
Page 106: ...106 Thermo Scientific Multidrop Combi nL User Manual Thermo Fisher Scientific Notes ...