Maintenance
Decontamination procedure
Thermo Fisher Scientific
Thermo Scientific KingFisher Duo User Manual
59
Perform decontamination in accordance with normal laboratory
procedures. Any decontamination instructions provided with the
reagents used should be followed.
It is strongly recommended to perform the complete decontamination
procedure before relocating the instrument from one laboratory to
another or before sending it to service.
Example of decontaminants:
●
Ethanol 70%
●
Virkon solution 1–3%
●
Glutaraldehyde solution 4%
●
Chloramine T
●
Microcide SQ 1:64
●
Decon 90 min. 4%
Warning
The decontamination procedure should be performed by
authorized trained personnel wearing disposable gloves, protective
glasses and clothing in a well-ventilated room.
▲
1.
Wear disposable gloves to protect yourself.
2.
Prepare the decontaminant: 200 ml 4% glutaraldehyde solution (or
another agent recommended by your safety officer).
3.
Empty the turntable.
4.
Switch OFF the power and disconnect the mains supply cable
(Figure 3–11).
5.
Disinfect the outside of the instrument using a cloth dampened
with 70% ethanol.
6.
Place the instrument in a large plastic bag. Ensure that the front lid
is open.
7.
Place a cloth soaked in the prepared solution into the bag. Ensure
that the cloth does not come into contact with the instrument.
8.
Seal the bag and leave the instrument in the bag for at least 24
hours.
Summary of Contents for KingFisher Duo
Page 1: ...Thermo Scientific KingFisher Duo User Manual Rev 1 0 ...
Page 2: ......
Page 3: ...Thermo Scientific KingFisher Duo User Manual Rev 1 0 Cat no N12420 ...
Page 12: ...Contents 10 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Page 76: ...System Log 74 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Page 78: ...Glossary 76 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Page 81: ...Thermo Fisher Scientific Thermo Scientific KingFisher Duo User Manual 79 Notes ...
Page 82: ...80 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific Notes ...
Page 83: ......