Routine Operation
How to start
54
Thermo Scientific KingFisher Duo User Manual
Thermo Fisher Scientific
functions with either one or two plates depending on the number of
steps. The tip comb always has to be placed manually onto a
KingFisher plate (Figure 4–29).
Only one tip comb
can be placed
onto a KingFisher plate per run
.
Figure 4–29.
Combining the tip comb into the B row of the plate
The two plate positions and the A1 positions of the two plate positions
are clearly marked on the turntable. When the instrument is in its basic
position, plate position 1 is under the KingFisher Duo head. Note that
at the end of a protocol run, the turntable may stop in a different
position than the starting position. Loading the plate and elution strip
is shown below (Figure 4–30 and Figure 4–31). Make sure that the
elution strip is placed in the correct direction into the elution block.
Ensure that the perforated end is facing towards the user.
Figure 4–30.
Loading the plate
A1 corner
Summary of Contents for KingFisher Duo
Page 1: ...Thermo Scientific KingFisher Duo User Manual Rev 1 0 ...
Page 2: ......
Page 3: ...Thermo Scientific KingFisher Duo User Manual Rev 1 0 Cat no N12420 ...
Page 12: ...Contents 10 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Page 76: ...System Log 74 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Page 78: ...Glossary 76 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific ...
Page 81: ...Thermo Fisher Scientific Thermo Scientific KingFisher Duo User Manual 79 Notes ...
Page 82: ...80 Thermo Scientific KingFisher Duo User Manual Thermo Fisher Scientific Notes ...
Page 83: ......