Installation
Operational check
Thermo Fisher Scientific
Thermo Scientific KingFisher Duo User Manual
29
4.
Take away the magnet head by turning the two finger screws on top
of the magnet head clockwise when instructed (Figure 3–15). When
you remove the KingFisher Duo head, be careful not to damage the
magnet rods against the tip comb holder frame.
Figure 3–15.
Removing the two finger screws on top of the magnet head
5.
Insert the other magnet head by slotting it into the tip comb holder
and by refitting the two finger screws on top of the magnet head
(Figure 3–16 and Figure 3–17).
6.
The tip moves on top of the shield plate at the end of the changing
process.
7.
Screw the maintenance door on again by screwing the two
red
screws on counterclockwise.
▲
Figure 3–16.
Inserting the KingFisher Duo head