Perform a manual cleaning of the sequencer
A cleaning protocol is normally performed automatically at the completion of each sequencing run. If a
cleaning is necessary, perform the following procedure.
1.
On the home screen, select
Settings
4
Clean Instrument
.
The instrument door unlocks allowing access to the consumables.
2.
Remove the Ion S5
™
ExT Wash Solution bottle to access the waste reservoir, then remove and
empty the waste reservoir.
1
2
3
3.
Reinstall the empty waste reservoir and a
used
Ion S5
™
ExT Wash Solution bottle.
4.
Ensure that the Ion S5
™
ExT Sequencing Reagents cartridge and Ion S5
™
ExT Wash Solution bottle
are properly installed.
IMPORTANT!
Perform the cleaning with a used reagent cartridge and wash solution bottle
installed. The cleaning procedure pumps cleaning solution into the wash solution bottle and
reagent cartridge making them unsuitable for sequencing.
B
Ion 520
™
& Ion 530
™
ExT Kit − Chef User Guide
65