Note:
Select
Pause
if you are uncertain of library quality and want to evaluate templating
efficiency before the chip is loaded. If you do not pause the run, you can collect QC samples
after the run. Save the samples until sequence analysis is complete to have them available for
troubleshooting.
10.
On the
Run Options
screen, tap
Start run
to start the run.
Note:
If you need to stop the run for any reason, tap
Cancel
, then tap
Yes
to confirm the
cancellation.
If the Ion Chef
™
Instrument encounters a problem during the run, it aborts the run and displays the
error on the instrument touchscreen. If a run fails, perform the following tasks.
a.
Remove the consumables from the deck, then clean the instrument. If possible, retain the
consumables for troubleshooting.
b.
Reset, then attempt to start the run again. If the run fails again, contact Technical Support to
troubleshoot the problem. Record the error message for reference.
11.
Initialize the sequencer at least 50 minutes before the Ion Chef
™
System finishes chip loading. See
Chapter 5, “Initialize the sequencer”.
12.
If you chose to pause the run to analyze the templating efficiency, remove the sample for testing
when prompted to do so by the Ion Chef
™
Instrument (approximately 5.75 hours into the run).
a.
When prompted to remove the QC sample, open the instrument door.
Chapter 4
Start the Ion Chef
™
run
Start the Ion Chef
™
run
4
40
Ion 520
™
& Ion 530
™
ExT Kit − Chef User Guide
Summary of Contents for Ion 520
Page 77: ......