77210163GB Issue 3
23
Note
1 Inadequate ventilation may have an adverse effect
on the cooling system. The recommended ambient
temperature is 20°C (68°F); maximum = 35°C
(95°F). Performance is adversely affected at ambient
temperatures above 30°C (86°F).
2 Siting of the instrument on an uneven surface can
have an adverse effect on the cooling system. Make
sure that the unit is level.
ThE EffICIENCY Of ThE ChAMBER CAN
BE AffECTED BY DOwNDRAUGhTS. DO
NOT SITE ThE INSTRUMENT NEAR TO
SOURCES Of DOwNDRAUGhTS SUCh AS
wINDOwS OR hEATING / VENTILATION
DUCTS
Turn the lowering knobs counter-clockwise until the
feet of the instrument are on the floor. The retraction
knobs can be unscrewed completely and removed
if preferred.
Raise the front edge of the top cover, remove the transit
packing from the top of the sliding window, then lower
the cover back into position. Push open the sliding
window to gain access to the Refrigerated Chamber.
Remove the rubber band and any other packing material
from the area around the Cryobar (h) on the left hand
wall of the Refrigerated Chamber.
h