Sections show ice
artifacts.
1 Specimen not frozen
quickly enough.
2 Specimen too thick.
1 Allow Cryoboost longer
time before adding
mountant and admitting
specimen.
2 Cut thinner.
Sections show signs of
the effects of vibration.
1 Knife not supported
correctly.
2 Knife angle incorrect.
3 Specimen not
sufficiently secured.
4 The blade is not
clamped securely
1 Disposable knives are most
vulnerable. Make sure that
the knife is fitted correctly
and that there is no debris
under the knife clamp.
2 Check and correct the knife
angle.
3 Re-freeze as necessary.
4 Make sure the blade
clamping is secure
Sections thick and thin.
1 Blunt knife
2 Knife angle too
shallow.
3 Specimen not secure.
4 Knife Holder is loose
1 Sharpen knife.
2 Widen the knife angle
3 Re-freeze and secure the
specimen.
4 Make sure the Knife Holder
is secure on the base
Sections stick to the Anti
Roll Plate.
1 Dirt or contaminant on
the Anti Roll Plate.
2 Cutting equipment not
cold enough.
1 Use the brush to clean the
Anti Roll Plate.
2 Allow more time for the
knife and the Anti Roll
Plate to cool.
Sections are split
vertically.
1 The knife is dirty.
2 The knife is blunt.
3 The knife/blade, anti-
roll plate or clamp plate
are damaged
1 Remove the knife and
clean and sharpen it.
wEAR MESh GLOVES.
2 Examine the knife/blade,
anti-roll plate and clamp
plate and replace if
necessary
TABLE 2 (continued)
PROBLEMS - whEN SECTIONING