77210163GB Issue 3
26
ThE CRYOTOME MUST BE PROTECTIVELY EARThED. MAKE SURE
ThAT ThE INSTRUMENT IS PLUGGED INTO A PROPERLY EARThED
(GROUNDED) MAINS SUPPLY
IT MUST BE POSSIBLE TO INTERRUPT ThE MAINS SUPPLY AT SOURCE
BY REMOVING ThE PLUG fROM ThE MAINS SUPPLY SOCKET
Select the appropriate mains lead from those supplied and plug
it into the mains inlet socket in the rear panel of the instrument
(b). Connect the mains supply cable to the local power supply
outlet.
MAKE SURE ThAT BODY Of ThE INSTRUMENT IS EffECTIVELY
EARThED (GROUNDED). A PROPRIETARY GROUND fAULT CIRCUIT
INTERRUPTION DEVICE (EARTh LEAK DETECTION) MAY BE fITTED AT
ThE wALL SOCKET AS AN ADDITIONAL SAfEGUARD
3.5 TO SwITCh ON AND Off
3.5.1
TO SwITCh ON
Press the l (ON)
side
of the
l / O switch inward to switch the
instrument on.
Check that the following occur during power on:
i
ALL lights and indicators of the Control Panel light for three seconds.
ii
E only: the Specimen Travel Available bar chart changes to one segment
flashing (when the automatic RESET function completes).
Standard only: the Specimen Travel Available bar chart remains flashing.
Rewind the Specimen Head to give maximum travel available (only one
segment flashing). Advance the Specimen Head to stop the single segment
flashing and to silence the alarm.