77210163GB Issue 3
17
Indications that show in the Time Controls display are as follows:
i
P fAIL (flashing) shows after a normal switch OFF then ON, or if power
failed and recovered. The Temperature display shows the temperature of
the Refrigerated Chamber at the time power was restored (i.e. the warmest
temperature).
Press
[CLOCK] to clear.
ii
rESEt (flashing) shows if a problem caused the instrument to reset and clear
its memory to its default status. The number in Temperature display is the
temperature of the Refrigerated Chamber at the time normal operation
resumed.
Press
[CLOCK] to clear.
iii
P SAVE shows until the programmed time if the instrument is in Standby
(PSAVE) mode following the selection of a DAY 9 Defrost or Fumigate function.
Press
[CANCEL DEfROST/fUMIGATE] to override.
iv
Err 1 denotes an invalid push-button entry, or if too many push-buttons are
pressed simultaneously.
Press
[CLOCK] to clear.
v
Err 2 denotes an invalid push-button entry during Defrost/Fumigate, or if Defrost
is inhibited because the Cryobar and Refrigerated Chamber are too warm
Press
[CLOCK] to clear.
vi
Err 4 denotes a window lock error (not locking or unlocking).
Press
[CLOCK] to clear.
Note
1 Err 3 is not used on these versions of Cryotome.